|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of gfbs1859 showed four other similar product complaint(s) from this lot number.The complaints for this lot number (gfbs1859) have been reported from the same facility.
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a broken/brittle stylet is confirmed for each of the 5 samples, but the cause is unknown.The devices returned were five navarre drainage catheters in their pouches, one of the pouches being opened (sample 1).In each case, the catheter, cap, and stylet were present in the pouch.In sample 1, the stylet was found to be broken in two pieces inside the open pouch.Both ends of the distal end of the broken stylet were fractured, and so it appears that some of the stylet tube is missing.No obvious damage was noted visually on the other devices.In samples 2-5, the stylet was curled into a circle, inserted fully into the catheter, and then 3) a thumb was run across the stylet distally while inducing a small angle in the stylet.In each of these samples, the stylet broke during this third step.In each sample, the stylet break(s) appeared well-defined and sharp, typical of brittle failure.The cause of this brittleness is unknown.This product will be sent to the manufacturing site for further review.A lot history review (lhr) of gfbs1859 showed four other similar product complaint(s) from this lot number.The complaints for this lot number (gfbs1859) have been reported from the same facility.
|