• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CODMAN & SHURTLEFF, INC. SURG PAT XRAY 1/2 X 3; PATTIE, COTTONOID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CODMAN & SHURTLEFF, INC. SURG PAT XRAY 1/2 X 3; PATTIE, COTTONOID Back to Search Results
Catalog Number 80-1407
Device Problem Detachment Of Device Component (1104)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Type  Injury  
Manufacturer Narrative
(b)(4) it has been communicated that the device and/or lot information is not available for evaluation.Without the device and/or lot information it is not possible for codman to conduct a proper investigation.If at some point the device and/or lot information does become available, this complaint will be re-opened, evaluated and a follow up report will be filed.Trends will be monitored for this and similar complaints.At the present time this complaint is considered closed.
 
Event Description
During a procedure,a the string broke off a pattie when it was being removed from the patient.The surgeon performed a brief unsuccessful search.Later, when the patient was x-rayed, the pattie was still not detected.Two days later the patient was returned to the or to retrieve the pattie.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURG PAT XRAY 1/2 X 3
Type of Device
PATTIE, COTTONOID
Manufacturer (Section D)
CODMAN & SHURTLEFF, INC.
325 paramount drive
raynham MA 02767
Manufacturer Contact
james kenney
325 paramount drive
raynham, MA 02767
5088282726
MDR Report Key7249256
MDR Text Key99302353
Report Number1226348-2018-10122
Device Sequence Number1
Product Code HBA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K880402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number80-1407
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/19/2018
Initial Date FDA Received02/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-