Catalog Number 1125350-18
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/01/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined that the reported failure to advance, stent dislodgement, removal of foreign body and subsequent treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the patient presented with chest pain and was intubated due to thrombosis of a previously non-abbott implanted stent in an ostium vein graft to the left anterior descending (lad) coronary artery.A 3.5 x 18 mm xience alpine stent delivery system (sds) was advanced but the distal end of the stent got caught with the proximal end of the pre-existing stent implant and the xience alpine stent dislodged.A snare device was used to retrieve the dislodged stent successfully.Dilatation was performed and a 3.0 x 28 mm xience alpine was used to complete the procedure successfully.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7249389
Report Number2024168-2018-00893
Device Sequence Number949516
Product Code NIQ
UDI-Device Identifier08717648199530
UDI-Public08717648199530
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/25/2020
Device Catalogue Number1125350-18
Device Lot Number7101241
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2018
Initial Report FDA Received Date02/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient Weight95
-
-