• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number DBP-200CLASS145
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 01/11/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported oad was received for analysis.When tested, the device was found to function as intended with no anomalies noted.Adhered biological material was observed on the driveshaft and crown, however no damage to the device was observed which would have contributed to the accumulation.At the conclusion of the failure analysis investigation, the cause of the reported event was unable to be confirmed.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
 
Event or Problem Description
During a peripheral atherectomy procedure using a csi orbital atherectomy device (oad), a dissection occurred.The target lesion was 90% stenosed and located in the common femoral artery.The lesion was treated using the oad on low speed, however when the device was operated on high speed it would continuously lose power.When the device would regain power, the crown would resume spinning.The device was replaced and treatment was continued.After use of the second oad, a dissection was noted.A stent was placed to resolve the dissection and the patient was stable following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul MN 55112
MDR Report Key7249456
Report Number3004742232-2018-00025
Device Sequence Number14028507
Product Code MCW
UDI-Device Identifier10852528005268
UDI-Public(01)10852528005268(17)190430(10)189218
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K133399
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/30/2019
Device Model NumberDBP-200CLASS145
Device Catalogue NumberDBP¿200CLASS145
Device Lot Number189218
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/11/2018
Initial Report FDA Received Date02/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-