ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896); Fail-Safe Problem (2936)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a call service alarm (communication) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the issue may result in long term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Additional Manufacturer Narrative
Date of submission 28-feb-2018 the device has been returned and evaluated by product analysis on 21-feb-2018 with the following findings: a review of the black box showed call service 052 alarms.The pump powered up with no alarms.The pump was run for 24 hours and no call service alarms were emitted.The call service alarm issue was unable to be duplicated during the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7249477
Report Number2531779-2018-02930
Device Sequence Number16433596
Product Code OYC
UDI-Device Identifier10840406100310
UDI-Public10840406100310
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2018
Was the Report Sent to FDA? Yes
Device Age10 MO
Initial Date Received by Manufacturer 01/24/2018
Supplement Date Received by Manufacturer01/24/2018
Initial Report FDA Received Date02/07/2018
Supplement Report FDA Received Date02/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age17 YR
-
-