• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MAPLE GROVE QUANTUM¿ MAVERICK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number H7493808015220
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The balloon was loosely folded with blood in the inflation lumen and the balloon.The outer shaft, inner shaft, balloon and tip were microscopically examined.The tip is damaged.The balloon has a pinhole 7mm from the proximal markerband.There are numerous hypotube and shaft kinks.Inspection of the remainder of the device presented no other damage or irregularities.There was no evidence of any material or manufacturing deficiencies contributing to the damage and the reported difficulty.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on 12-jan-2018.It was reported that crossing difficulties were encountered.The target lesion was located in the left anterior descending artery.A 2.25mm x 15mm quantum¿ maverick¿ balloon catheter was advanced for dilation but device failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed balloon pinhole.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7250056
Report Number2134265-2018-00431
Device Sequence Number1483963
Product Code LOX
UDI-Device Identifier08714729392750
UDI-Public08714729392750
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/21/2020
Device Model NumberH7493808015220
Device Catalogue Number38080-1522
Device Lot Number0020804636
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2018
Initial Report FDA Received Date02/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-