| Model Number |
N/A |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of 17jb2529 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
The facility reported to the sales rep that during priming the picc with saline prior to placement, vascular access nurse noticed saline appeared to exit the catheter from a site other than the catheter tip.It was stated the nurse disconnected syringe and reconnected to assure all connections were correct and noticed a vertical crack in the catheter hub.On attempt to show staff where catheter was cracked it was reported that the catheter hub split at the colored bifurcation.Another kit was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a crack in the gray connector was confirmed, and was considered manufacturing related.The implicated and returned product was a 6fr triple-lumen powerpicc solo.The length of the t/l tubing had not been modified.The gray connector was broken and received in two segments.The proximal half of the connector was attached to a syringe.A longitudinal crack was observed in the proximal half of the connector.The valve from the gray connector was received over the t/l tubing.Cavity ¿j¿ was identified in the gray connector.No cracks were observed on the other connectors.A microscopic examination revealed that the crack had sharp edges and no plastic deformation was observed.Additional longitudinal cracks in the gray connector were observed with the microscope.One crack was observed on what appeared to be the knit/weld line.The complaint sample was forwarded to the manufacturing facility for further review.Reynosa evaluation: the complaint for ¿vertical crack in the catheter hub¿ is confirmed according the evaluation the sample received which showed a microscopic evaluation showed a break in the molding valve, maybe those are caused by the incorrect molding process, leading to valve breakage.This condition cause leak in the device.Therefore the cause for this complaint is manufacturing related.A lot history review (lhr) of 17jb2529 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
The facility reported to the sales rep that during priming the picc with saline prior to placement, vascular access nurse noticed saline appeared to exit the catheter from a site other than the catheter tip.It was stated the nurse disconnected syringe and reconnected to assure all connections were correct and noticed a vertical crack in the catheter hub.On attempt to show staff where catheter was cracked it was reported that the catheter hub split at the colored bifurcation.Another kit was used to complete the procedure.Additional details: prior to placement of picc, noticed line had cracked hub.Crack noticed by flushing with saline.Crack noticed when syringe connected to hub and attempted to flush, squirting saline in face.Upon closer inspection, crack was noticed running perpendicular to the opening of the hub when syringe tip screwed onto hub.Defect noticed prior to inserting patient.
|
| |
|