| Model Number |
97000HS |
| Medical Device Problem Code |
Incorrect Or Inadequate Test Results (2456)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation pending.Malfunction mdr for lot w63243b report number: 3013982035-2018-00004.
|
| |
|
Event or Problem Description
|
|
The customer reported observing a variance for myoglobin during a lot to lot comparison with two different triage cardiac lot numbers.Patient results: triage cardiac lot w63243b, myo=90.1, triage cardiac lot w63316b, myo=118.Myoglobin reference range: 0-106 ng/ml.Patient was not treated based on the triage results.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation conclusion: the customers complaint was not replicated with in-house testing of retain lot w63316b.No issues with myo recovery observed.Manufacturing batch records for lot w63316b were reviewed and found that the lot met final release specifications.Based on the information available, there is no indication of a product deficiency and no corrective action is required.
|
| |
|