ENCORE MEDICAL L.P. FOUNDATION HIP; SHELL, FMP HEMISPHERICAL SIZE54 W/HOLES Back to Search Results
Model Number 430-01-054
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/29/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to a malpositioned cup.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to malposition of cup.The previous surgery and the revision detailed in this investigation occurred 2.9 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to malposition of cup.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: patient activities, inadequate soft tissue support, excessive range of motion, excessive loads or torques, poor bone quality, improper surgical procedure or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION HIP
Common Device Name
SHELL, FMP HEMISPHERICAL SIZE54 W/HOLES
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7251013
Report Number1644408-2018-00106
Device Sequence Number140938
Product Code LPH
UDI-Device Identifier00888912081832
UDI-Public(01)00888912081832
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K974093
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2020
Device Model Number430-01-054
Device Catalogue Number430-01-054
Device Lot Number285C1087
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/29/2018
Supplement Date Received by Manufacturer03/15/2018
Initial Report FDA Received Date02/08/2018
Supplement Report FDA Received Date03/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
010-55-035, LOT 009A1106; 400-03-361, LOT 864B1246; 931-36-254, LOT 692F1128; 010-55-035, LOT 009A1106; 400-03-361, LOT 864B1246; 931-36-254, LOT 692F1128
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
-
-