It was reported that during a fenestrated graft procedure in the abdominal aorta, the 14 fr performer introducer was leaking at the valve near/ while a wire guide was in the middle of the valve.The physician stated that this issue has occurred on previous occasions during the last several months (refer to medwatch 1820334-2018-00392); an eight (8) fr sheath was/ is placed inside of the lumen of the 14 fr introducer to stop the leakage.The patient lost a lot of blood (unknown quantity) during the procedure; however, a blood transfusion was not required.The graft was initially unable to be placed; however, the physician re-attempted and the procedure was able to be successfully completed.A section of the device did not remain inside the patient¿s body.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation ¿ evaluation: a review of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
|