| Model Number |
H7493952838300 |
| Medical Device Problem Code |
Component Incompatible (1108)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.Device evaluated by mfr.: the promus premier us mr 3.00 x 38 mm stent delivery system was returned for analysis.The stent was detached from the sds and was not returned for analysis.There were no issues reported with the stent.The balloon wings were relaxed and it appeared the balloon had been inflated and a vacuum pulled.There was hardened contrast medium evident in the balloon body.The device was attached to a handheld encore inflation device without issue.The balloon was inflated to nominal inflation pressure without issue.A visual and tactile examination of the device found multiple hypotube kinks.An examination of the shaft polymer extrusion found no issues.The bi-component bond showed no signs of damage or strain.The tip was visually examined and no issues were noted.No other issues were identified during the product analysis.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 15-jan-2018.It was reported that connection incompatibility occurred.A 3.00 x 38 mm promus premier¿ drug-eluting stent was selected for use.However, when the stent delivery system (sds) was connected to a bsc inflation device, the sds would not connect securely.The procedure was completed with another 3.0 x 38 mm promus drug-eluting stent.No patient complications were reported.However, returned device analysis revealed a detached/separated stent.
|
| |
|
Search Alerts/Recalls
|