U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY® 3; CATHETER,INTRAVASCULAR,THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY® 3; CATHETER,INTRAVASCULAR,THE Back to Search Results
Model Number 4251130-02
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 01/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2016018 b.Braun inc.(bbmi) (importer) is submitting the report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.A follow-up report will be submitted when the results of the investigation are available.
 
Event or Problem Description
As reported by user facility: the patient was wincing when trying to place the catheter.The catheter was removed and it was then noted that the catheter had sheared in the middle.The patient was restuck without further incident.
 
Additional Manufacturer Narrative
Exemption number e2016018.This report has been identified as b.Braun inc.Internal report number (b)(4).Two used catheters were received for evaluation.When the two samples were visually inspected, one sample had a pierced catheter about 12mm from the needle tip, while the other sample had no evidence of damage.This product is subjected to a 100% inspection by an in-line vision system that is calibrated and subjected to regular verifications to ensure it is functioning properly.The defect observed in the returned sample would have been detected and rejected by the in-line vision system.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number.Based on the results of this investigation, no specific conclusions can be made regarding the cause of the reported event.Although no conclusion can be made, the most likely cause of the reported event is that the needle was reinserted.The instructions for use (ifu) clearly states, "after withdrawal, do not reintroduce the steel needle into the catheter, as the latter may be cut off".If additional pertinent information becomes available a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY® 3
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key7251357
Report Number9610825-2018-00003
Device Sequence Number1215765
Product Code FOZ
UDI-Device Identifier04046964014055
UDI-Public(01)04046964014055
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K111236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 02/16/2018
Report Date (Section F) 02/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/01/2022
Device Model Number4251130-02
Device Catalogue Number4251130-02
Device Lot Number17H15G8373
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2017
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/10/2018
Device Age5 MO
Date Report to Manufacturer02/16/2018
Initial Date Received by Manufacturer 01/10/2018
Supplement Date Received by Manufacturer01/10/2018
Initial Report FDA Received Date02/08/2018
Supplement Report FDA Received Date02/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
Date Report Sent to FDA02/16/2018
-
-