| Catalog Number |
750190 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Airway Obstruction (1699)
|
| Date of Event |
12/14/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Information was received that a patient with a smiths medical bivona® adult aire-cuf® tracheostomy tube was not receiving air.Subsequently, the tube was changed out and the patient was able to breathe normally.No further adverse effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was returned for evaluation.Visual inspection of the device did not identify any defects.Leak testing was performed with air and did not identify any leaks.An in-house clinician evaluated that patient anatomy may have caused the device to be over-filled to lead to the reported issue.Based on the evidence, a root cause was unable to be confirmed.No fault was found with the device.
|
| |
|