NEODENT - JJGC S.A. CM TITAMAX EX ACQUA IMPLANT 4.0X17; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.640
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Unspecified Infection (1930); Tissue Damage (2104); Sinus Perforation (2277)
Date of Event 10/03/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.Instradent usa, inc is submitting the report on behalf of neodent - jjgc.
 
Event or Problem Description
(b)(4).The dentist reported that 20 days after dental implant was installed in mouth, its non-osseointegration was observed.A 35 ncm of primary stability was achieved and the implant was installed with immediately load.The patient presented bone type iv.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM TITAMAX EX ACQUA IMPLANT 4.0X17
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7251536
Report Number3008261720-2018-00629
Device Sequence Number16482653
Product Code DZE
UDI-Device Identifier07899878016595
UDI-Public07899878016595
Combination Product (Y/N)N
Initial Reporter CountryES
PMA/510(K) Number
K133592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 02/08/2018
Report Date (Section F) 01/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Expiration Date11/30/2018
Device Catalogue Number140.640
Device Lot Number85840
Other Device ID Number07899878016595
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/08/2018
Device Age11 MO
Event Location Other
Date Report to Manufacturer01/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2018
Initial Report FDA Received Date02/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-