| Model Number |
THE BAG II |
| Medical Device Problem Codes |
Ambient Temperature Problem (2878); Device Operational Issue (2914); Activation Failure (3270)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The user's temperature tested resuscitator has not been returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
The prehospital service department of a non-us hospital reported the bag ii resuscitator's ability to re-inflate starts to slow at -7 degrees celsius and by -18 c (dfu, minimum operating temp) the ventilation bag does not re-inflate at all; as found during their in-house temperature testing.They reported they have previously submitted several nonconformance reports to their local health authorities for other patient incidents (no details provided) but during these temperature tests no patient was involved nor harmed.
|
| |
|