WRIGHTS LANE: SYNTHES USA PRODUCTS LLC 11MM/130 DEG TI CANN TFNA 420MM/RIGHT - STERILE; ROD,FIXATION,INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
04.037.162S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient information was not provided for reporting.Date of event is unknown.Date of implant is unknown.Device is expected to be returned for manufacturer review/investigation, but has not been received yet.(b)(4).Device evaluation/investigation could not be completed; no conclusion could be drawn as product was not returned to manufacturer.Device history records review could not be completed without lot number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, patient underwent a hip surgery.Approximately six months post- operative, during the examination, surgeon noted the absence of callus formation and that the implanted long tfna nail was broken under stress at the nail/screw interface.Due to the implant failure, the nail was removed on (b)(6) 2018 which was difficult given where the implant failed.Once removed, a larger diameter nail was inserted.Patient outcome was not reported.Concomitant devices reported: lag screw (part# unknown, lot# unknown, quantity 1).This report is for one (1) 11 mm/130 deg ti cann tfna 420 mm/right ¿ sterile.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was further reported that the nail broke at the nail/screw junction which made it difficult to remove as there are no threads or anything to hook into.Therefore, the surgeon used a burr to create a hole for a hook to fit into the nail to remove.The lag screw did not break and there was no malfunction associated with the original lag screw or the lag screw used in the revision surgery.There was approximately 30-45 minutes delay in procedure.Revision surgery was successfully completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Dhr review was completed.Manufacturing location: monument.Manufacturing date: 11-dec-2014.Expiration date: 31-oct-2024.Part #: 04.037.162s, lot#: 7853156 (sterile) - 11mm/130 deg ti cann tfna 420mm / right - sterile.Quantity 6.Component parts reviewed: part 04.037.942.2 - lock prong, 130 degree, tfna bp-55 lot - 9143928.Part 04.037.912.4 - wave spring, shim ended bp-55 lot - 7722144.Part 04.037.912.3 - tfna lock drive bp-58 lot - 7754636.Part 21127 - raw material lot bp-80 lot ¿ 7664309.Raw material for ti received from perryman company.Certified test report received from perryman and certificate of test received for titanium ingot meet specification.Raw material receiving/putaway checklist meet requirements.Inspection sheet for in-process/inspect dimensional/final met inspection acceptance criteria.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.¿sterility documentation was reviewed and determined to be conforming.¿ customer quality conducted an investigation of the returned device.The 11mm/130 degree 420mm / right titanium cannulated tfna nail (04.037.162s) is an implant in the titanium tfna system for intramedullary fixation of proximal femur fractures.The returned device was examined at customer quality (cq) and the complaint condition was able to be confirmed as the device was found to be broken at the proximal locking hole.The break profile is oblique and jagged.Both pieces were returned.No root cause could be determined with the provided information.The complaint condition was unable to be replicated due to post-manufacturing damage.Relevant drawings for the returned device were reviewed (both current revision and from the time of manufacture): top-level.The design, materials and finishing processes were found to be appropriate for the intended use of this device.No dimensional analysis is possible due to post-manufacturing damage (break profile is oblique and jagged).A device history review, including material review was performed for the returned implant¿s lot number and no mrrs, ncrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.The tfna risk document adequately addresses the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|