| Model Number |
DVBC3D1 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) was attempted to be implanted but not used due to a setscrew problem.The setscrew was unable to be seated into threads.The icd was removed from implant and a new icd was implanted.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.The returned device had foreign material in the connector and a damaged connector.The returned device indicated a grommet issue, a damaged grommet, foreign material on set screw, a set screw with a rounded socket, and a damaged set screw.If information is provided in the future, a supplemental report will be issued.
|
| |
|