• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO SECURA DR; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO SECURA DR; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number D204DRM
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant product(s): a 5076-52 lead, implanted on (b)(6) 2013.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient¿s implantable cardioverter defibrillator (icd) experienced premature battery depletion with device longevity of less than five years.It was also reported that the patient¿s right ventricular (rv) lead had elevated thresholds.The icd was explanted and replaced.The rv lead remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product analysis the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECURA DR
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7254236
Report Number3004209178-2018-02496
Device Sequence Number16170750
Product Code LWS
UDI-Device Identifier00613994714695
UDI-Public00613994714695
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/28/2014
Device Model NumberD204DRM
Device Catalogue NumberD204DRM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2018
Initial Date Received by Manufacturer 12/28/2017
Supplement Date Received by Manufacturer02/20/2018
Initial Report FDA Received Date02/08/2018
Supplement Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
-
-