| Model Number |
693558 |
| Medical Device Problem Codes |
Failure to Capture (1081); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/07/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned.Analysis was performed and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
About three weeks post implant, that the right ventricular (rv) lead had pulled back and had no capture at five volts.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|