| Catalog Number |
500-55150 |
| Medical Device Problem Code |
Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to ekos and evaluated on 10 jan 2018.Evaluation of the returned msd revealed evidence of shorted wires.The iddc was in used condition but no damage was observed.Evaluation conclusion - the root cause of the shorted wires was not identified.A review of the device history record for the msd confirmed that the product was manufactured per standard processes and met all acceptance criteria.No patient impact was reported.The damage did not affect the iddc and therefore did not come into contact with the patient's vasculature.The event did not impact the delivery of thrombolytic therapy.No serious injury occurred as a result of this malfunction.The control unit event log recorded that ultrasound was delivered for a total of 24.9 hours.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, a helpline call reported a temperature alarm after 16 hours and 40 minutes of therapy.The alarm was resolved over the phone with the ekos representative by increase the coolant rate through the device.The device was returned to ekos for evaluation.
|
| |
|
Search Alerts/Recalls
|