| Model Number |
5019 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/03/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated pacing capture threshold in the right ventricle was elevated.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported the patient required placement of an additional array due to high defibrillation testing where three shocks were needed to convert the rhythm.The array was connected via the adaptor to the device.Defibrillation testing was performed with success and all the measurements were found to be within normal range.Approximately two or three days later, the right ventricular lead displayed high thresholds, there was a jump in impedance and the impedance was high.Testing revealed the high values were repeatedly confirmed.Following, the patient was admitted to the operating room for a lead revision; however, at that time the measurements were found to be normal.It was also reported there was a possibility there was an interaction with the adaptor.The cardiologist and surgeon decided to follow the patient and the lead revision was not performed.The device and lead remain in use and no patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|