| Model Number |
407652 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis summary: the full lead was returned, analyzed, and no anomalies were found.The distal low voltage electrode of the lead was covered in body tissue/fibrotic growth.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during implant of the right ventricular (rv) lead, the lead exhibited noise and undersensing.The lead was removed an an alternative lead was implanted.No patient complications have been reported as a result of this event.
|
| |
|