Model Number H74939293061530
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device evaluated by manufacturer: the device was not returned for analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
It was reported that the a stent placement problem occurred.The target lesion was a dissection which originated from an eccentric smooth lesion in the minimally tortuous right superficial femoral artery (sfa).A contralateral approach was made to access the lesion and a 6x150x130 innova¿ stent was advanced over a.035¿ guidewire.The stent was deployed with no mechanical issues noted; however, a portion of the proximal end of the stent ended up in the common femoral across the profunda.The stent was not stretched or elongated.The patient did not receive additional treatment and was monitored overnight.No patient complications occurred and the patient¿s status is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA¿
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7255241
Report Number2134265-2018-00733
Device Sequence Number1066311
Product Code NIP
UDI-Device Identifier08714729874041
UDI-Public08714729874041
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P140028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberH74939293061530
Device Catalogue Number39293-06153
Device Lot Number19887588
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2018
Initial Report FDA Received Date02/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
-
-