MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO PROTECTA CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO PROTECTA CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number D334TRM
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/19/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) had premature battery depletion.The device was explanted and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary : the device was returned and analyzed.The actual longevity was less than 80% of the 99.9% longevity limit.The device was fully functional, with no high current drain or evidence of battery problems.Without the history of the programmed settings throughout its service life, there is no way to determine why the longevity did not match the predicted model.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA CRT-D
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7255541
Report Number3004209178-2018-02647
Device Sequence Number1066843
Product Code NIK
UDI-Device Identifier00613994714619
UDI-Public00613994714619
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P010031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/14/2014
Device Model NumberD334TRM
Device Catalogue NumberD334TRM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2018
Initial Date Received by Manufacturer 12/19/2017
Supplement Date Received by Manufacturer02/20/2018
10/13/2019
Initial Report FDA Received Date02/08/2018
Supplement Report FDA Received Date04/10/2018
10/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
-
-