| Model Number |
407652 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/09/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned, analyzed, and no anomalies were found.The distal low voltage electrode of the lead was covered in body tissue/fibrotic growth.The outer insulation of the lead was extrinsically breached due to a cut.Visual analysis of the lead indicated damage at implant.Visual analysis of the lead indicated the lead was stretched.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during post implant, the right atrial (ra) lead dislodged.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|