| Model Number |
5071-25 |
| Medical Device Problem Codes |
Low impedance (2285); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right ventricular (rv) lead and right atrial (ra) lead exhibited high thresholds.It was further noted that the implantable pulse generator (ipg) battery longevity estimator measurement was not available, however, the battery voltage was available.The device and leads remain in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported low impedance was noted on the ra lead and rv leads.The longevity estimate was low on the ipg.The ipg and leads remains in use.
|
| |
|