| Model Number |
6248V-90P |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/14/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary : the delivery catheter was returned and analyzed.The mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.The mechanical operation of the catheter was damaged.Visual analysis of the catheter indicated damage during use.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during implant, the catheter broke into two pieces while slitting.The catheter was not used and replaced.No patient complications have been reported as a result of this event.
|
| |
|