• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TYRX, INC. AIGIS-R LARGE; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TYRX, INC. AIGIS-R LARGE; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number CMRM6133
Medical Device Problem Code Device Expiration Issue (1216)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/03/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the antibacterial envelope was placed and before the pocket was closed it was realized the envelope was past the expiration date by three days.The envelope was removed and a new one implanted.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIGIS-R LARGE
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
TYRX, INC.
1 deer park dr.
suites g, e
monmouth jct NJ 08852
MDR Report Key7257422
Report Number3005619263-2018-00023
Device Sequence Number894427
Product Code FTL
UDI-Device Identifier00643169468795
UDI-Public00643169468795
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K130943
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/31/2017
Device Model NumberCMRM6133
Device Catalogue NumberCMRM6133
Device Lot Number17H25157
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/03/2017
Supplement Date Received by Manufacturer10/28/2018
Initial Report FDA Received Date02/08/2018
Supplement Report FDA Received Date11/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
-
-