| Model Number |
C304L6906 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/30/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during implant, the physician noticed the braidwire exposed post slitting of the catheter.The catheter was removed.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the partial of the delivery catheter was returned and analyzed.Analysis indicated that the mechanical operation wire was pulled and exposed through the insulator.The catheter shaft was kinked/buckled.The mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.Visual analysis of the lead indicated damage during use.If information is provided in the future, a supplemental report will be issued.
|
| |
|