Model Number ZXT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt150, 11.5 diopter intraocular lens (iol) was explanted as the patient was unable to tolerate the multifocal lens.The lens was exchanged for a regular toric iol, model zct150 (13.0 diopter).There was no incision enlargement and no sutures required.There was no patient post-op injury reported.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 2/16/2018.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint event is not confirmed.Manufacturing record review: manufacturing records were reviewed and the device was manufactured according to specification.A search on complaints history revealed no similar complaints were received for this production order number to date.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key7258513
Report Number9614546-2018-00131
Device Sequence Number1158764
Product Code POE
UDI-Device Identifier05050474605039
UDI-Public(01)05050474605039(17)220225
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date02/25/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Initial Date Received by Manufacturer 01/12/2018
Supplement Date Received by Manufacturer04/09/2018
Initial Report FDA Received Date02/08/2018
Supplement Report FDA Received Date05/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-