| Model Number |
ZXT150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxt150, 11.5 diopter intraocular lens (iol) was explanted as the patient was unable to tolerate the multifocal lens.The lens was exchanged for a regular toric iol, model zct150 (13.0 diopter).There was no incision enlargement and no sutures required.There was no patient post-op injury reported.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 2/16/2018.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint event is not confirmed.Manufacturing record review: manufacturing records were reviewed and the device was manufactured according to specification.A search on complaints history revealed no similar complaints were received for this production order number to date.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|