| Catalog Number |
75446545 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
X-ray review result: post-op x-ray for l4-5 tlif showed a detachment of one of the tulip heads from the screw.Fusion status is unknown.This appears at the superior screw.This may be caused by improper final tightening/torque/counter torque.Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an unspecified spinal surgery due to l4-l5 lysis.Post-op, during a follow up visit the implanted screw head was found to be detached from screw tail.No patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Procedure performed: minimally invasive spinal fusion no additional/revision surgery has been performed to explant the product.Patient condition is stable.
|
| |
|