MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY PROMUS PREMIER¿; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY PROMUS PREMIER¿; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number H7493952828220
Medical Device Problem Codes Migration or Expulsion of Device (1395); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/26/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device is a combination product.(b)(4).It is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.(b)(4).
 
Event or Problem Description
It was reported that a stent dislodged and was surgically removed.The target lesion was a calcified left anterior descending artery.A 2.25 x 28mm promus premier¿ stent was advanced and had difficulty crossing the lesion.A little force was used and then the balloon partially came off the balloon.The stent delivery system and stent were still on the guidewire; the physician intended to pull the entire system with the guidewire into the guide catheter and remove everything from the body.Upon doing this, the stent became further dislodged and embolized.The patient was x-rayed and the stent had migrated to the patient's ankle.The patient was transported to the interventional radiology lab where a vascular surgeon removed the stent.There were no further patient complications and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER¿
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7259162
Report Number2134265-2018-00604
Device Sequence Number1022825
Product Code NIQ
UDI-Device Identifier08714729844884
UDI-Public08714729844884
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P110010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/17/2019
Device Model NumberH7493952828220
Device Catalogue Number39528-2822
Device Lot Number21302850
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/17/2018
Initial Report FDA Received Date02/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-