• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY MAILLEFER X-SMART ENDO MOTOR WITH CONTRA ANGLE HANPIECE HANDPIECE, DIRECT DRIVE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY MAILLEFER X-SMART ENDO MOTOR WITH CONTRA ANGLE HANPIECE HANDPIECE, DIRECT DRIVE, AC-POWERED Back to Search Results
Catalog Number A100400000000
Device Problems Retraction Problem (1536); Self-Activation or Keying (1557); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury. Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such. As such, this event is reportable per 21cfr part 803. The device is available for evaluation, though has not been returned as of this report. Evaluation results will be submitted as they become available.
 
Event Description
In this event it was reported that a xsmart suddenly goes into auto reverse mode; no injury resulted.
 
Manufacturer Narrative
Evaluation found some rust on the head cap and the cartridge is rusted, both are due to improper maintenance by the user. Also, the rotation is not fluid in the neck.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameX-SMART ENDO MOTOR WITH CONTRA ANGLE HANPIECE
Type of DeviceHANDPIECE, DIRECT DRIVE, AC-POWERED
Manufacturer (Section D)
DENTSPLY MAILLEFER
chemin du verger 3
ballaigues, 1338
SZ 1338
Manufacturer (Section G)
NAKANISHI INC.
700 shimohinata
kanuma-shi, 322-8 666
JA 322-8666
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7259448
MDR Text Key100003161
Report Number8031010-2018-00022
Device Sequence Number1
Product Code EKX
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K990682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 04/04/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberA100400000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received03/06/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-