• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ BI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D132702
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 01/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was discarded, therefore no product failure analysis can be conducted and device malfunction cannot be confirmed.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
 
Event or Problem Description
It was reported that a patient underwent a pulmonary vein isolation (pvi) ablation procedure for atrial fibrillation with a thermocool smarttouch bidirectional sf catheter and suffered a cardiac tamponade requiring pericardiocentesis.Pre-procedure, a small pericardial effusion was noted via intracardiac echocardiography (ice).After transseptal puncture, heparin administration, and extensive mapping, to include difficult access to the left superior pulmonary vein (lspv), the pericardial effusion volume was noted to have increased.Pericardiocentesis was performed and yielded 140 ml of fluid.Remainder of procedure was aborted.Patient remained stable throughout.No ablation was performed.It was noted that the physician does not believe that a bwi product was responsible for the injury.Multiple attempts have been made to obtain clarification to this complaint.However, no further information has been made available.
 
Additional Manufacturer Narrative
Additional information was received on february 7, 2018.The patient involved was female.The patient did not require extended hospitalization due to the event.Since the event the patient has fully recovered.The physician did not provide a causality opinion on the cause of the adverse event.However, it was noted that the patient had small pulmonary veins, which may have contributed to the event.A transseptal puncture was performed.The patient received heparin as an anticoagulant after transseptal puncture and the activated clotting time was maintained at 300-350 seconds.A st.Jude medical sl1 8fr sheath was used during the procedure.No radiofrequency was delivered during the procedure.It is unknown when the injury occurred.The flow rate was 2ml/min.There were no error messages observed on biosense webster equipment during the procedure.The smarttouch catheter was not in close proximity to another catheter.The catheter was not zeroed after the initial warm-up phase post catheter connection to the carto® 3 patient interface unity.The ablation catheter had not been used in procedure; it was introduced but not used.Manufacturer's ref.No: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH¿ BI-DIRECTIONAL NAVIGATION CATHETER
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
33 technology drive
irvine CA 92618
MDR Report Key7259603
Report Number2029046-2018-01174
Device Sequence Number13749460
Product Code LPB
UDI-Device Identifier10846835009170
UDI-Public10846835009170
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030031/S053
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2018
Device Catalogue NumberD132702
Device Lot Number17755049M
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/18/2018
Supplement Date Received by Manufacturer01/18/2018
Initial Report FDA Received Date02/09/2018
Supplement Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
-
-