• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERPICC SOLO2 CATHETER 4F MAXIMAL BARRIER TRAY WITH MICROINTRODUCER (S/L); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERPICC SOLO2 CATHETER 4F MAXIMAL BARRIER TRAY WITH MICROINTRODUCER (S/L); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a broken stylet is inconclusive due to poor sample condition.One photo sample of a stylet was returned for evaluation.An initial visual observation of the photo showed what appears to be a portion of a tls stylet.Bends in the stylet are visible through the portion of the stylet showed in the photo provided.The corresponding portion of the broken stylet is not shown.Based on the description of the reported event and photo sample provided, possible contributing causes include damage to stylet during catheter trimming and kink stylet damage.The photo does show a portion of the stylet but it does not show what device it originated from; consequentially, the complaint of a broken stylet is inconclusive due to poor sample condition.No further action is required because the reported event could not be related to a deficiency in product manufacture or deficiency while under correct product use.A lot history review (lhr) of rebx2245 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that during insertion of a picc, the guidewire broke and was found in the subclavian vein.The patient was taken to surgery and the wire was removed by a vascular surgeon.It was stated the wire was about 3¿ in length.It was later reported that the nurse thought it may be from an old picc placement.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a break in the tls stylet was confirmed, and the damage appeared to be related to use of the device.The distal 4.3cm segment of tls stylet was returned for investigation.The returned sample resembled the item that was in the provided photograph.The polyimide tubing was bent 1.7cm from the distal tip.The proximal segment of the stylet was not returned for investigation.The proximal end of a magnet was protruding less than ½ mm from the proximal end of the polyimide tubing.Kinks were observed in the polyimide tubing between magnets.This type of damage can occur when the polyimide tubing is kinked and flexed against the magnets.The powerpicc solo ifu states, ¿avoid sharp or acute angles during implantation.¿ the ifu for the 3cg stylet states, ¿ensure that the stylet tip does not extend beyond the trimmed end of the catheter.Extension of the stylet tip beyond the catheter end, combined with kinking and excessive forces, may result in vessel damage, stylet damage, difficult removal, stylet tip separation, potential embolism and risk of patient injury.¿ ¿avoid placement or securement of the catheter where kinking may occur, to minimize stress on the catheter, patency problems or patient discomfort.¿ no further action is required because the reported event could not be related to a deficiency in product manufacture or deficiency while under correct product use.A lot history review (lhr) of rebx2245 showed no other similar product complaint(s) from this lot number.(b)(4).
 
Event or Problem Description
It was reported that during insertion of a picc, the guidewire broke and was found in the subclavian vein.The patient was taken to surgery and the wire was removed by a vascular surgeon.It was stated the wire was about 3¿ in length.It was later reported that the nurse thought it may be from an old picc placement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO2 CATHETER 4F MAXIMAL BARRIER TRAY WITH MICROINTRODUCER (S/L)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7259639
Report Number3006260740-2018-00167
Device Sequence Number16433793
Product Code LJS
UDI-Device Identifier00801741035609
UDI-Public(01)00801741035609
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K072230
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 04/30/2018
Report Date (Section F) 01/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number9194108D
Device Lot NumberREBX2245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/13/2018
Event Location Hospital
Date Report to Manufacturer02/26/2018
Initial Date Received by Manufacturer 01/24/2018
Supplement Date Received by Manufacturer03/05/2018
Initial Report FDA Received Date02/09/2018
Supplement Report FDA Received Date04/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/23/2018
-
-