• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383400
Device Problem Delayed Alarm (1011)
Patient Problem No Patient Involvement (2645)
Event Date 11/07/2017
Event Type  malfunction  
Manufacturer Narrative
The customer complaint that the unit failed the rotor error alarm section of the performance verification was confirmed.A formal corrective and preventative action investigation has been opened to address the issue.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer states that the unit failed the rotor error alarm test and alarmed for feed error instead.Service found that the unit failed the rotor error test alarming for flow error after 2 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
Manufacturer (Section G)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7259724
MDR Text Key99919679
Report Number3008361498-2018-00073
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 02/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2017
Initial Date Manufacturer Received 11/08/2017
Initial Date FDA Received02/09/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-