| Model Number |
U4200210RX |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The results of the anticipated device evaluation will be provided upon completion of the event investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure of a calcified lesion in the tibial artery, a break was allegedly identified on the catheter shaft of the pta balloon device.The balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.The device was returned for evaluation.A visual inspection found a complete detachment in the catheter shaft, confirming the investigation for the reported detachment.It is likely the user perceived the detachment as a break.The definitive root cause for the identified detachment could not be determined based upon available information.It is unknown whether procedural issues contributed to the event.Labeling review: the current ifu (instructions for use) states: warnings: when the catheter is exposed to the vascular system, it should be manipulated while under high quality fluoroscopic observation.Do not advance or retract the catheter unless the balloon is fully deflated.If resistance is met during manipulation, determine the cause of the resistance before proceeding.Applying excessive force to the catheter can result in tip or catheter breakage, catheter kink, or balloon separation.
|
| |
|
Event or Problem Description
|
|
It was reported that during an angioplasty procedure of a calcified lesion in the tibial artery, a break was allegedly identified on the catheter shaft of the pta balloon device.The balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
|
| |
|