• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERPICC SOLO CATHETER WITH SHERLOCK 3CG TPS STYLET 5F BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERPICC SOLO CATHETER WITH SHERLOCK 3CG TPS STYLET 5F BASIC TRAY; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Health Effect - Clinical Codes Skin Erosion (2075); Thrombus (2101)
Date of Event 12/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of clotting and skin breakdown due to an inserted picc cannot be confirmed due to the condition of the sample, as well as unknown clinical conditions.The sample returned consisted of four photographs, apparently of a patient.Visual observation found that the photographs showed a lesion of some kind, wherein redness and the appearance of a bruise or similar was evident.The skin appeared wrinkled in the area of this lesion and some of the top layer of skin appeared to have come off.No catheter was present in any of the photographs.One of the photographs showed an area of skin which was apparently not affected like the area of skin shown in the other photographs.However, blood vessels were visible and appeared to be raised.Although a lesion was found depicted in the returned photographs, no link can be objectively determined between the catheter and this patient condition with the photographs and evidence available.The complaint therefore cannot be confirmed.No further action is required because the reported event could not be related to a deficiency in product manufacture or deficiency while under correct product use.The ifu states the following: "possible complications: ¿hematoma: hypersensitivity, anaphylactic or anaphylactic-like reaction during placement1, positioning1, flushing2 of catheter or cleaning of catheter exit site3.Intolerance reaction to implanted device perforation of vessels or viscus.Phlebitis.Venous thrombosis: risks normally associated with local or general anesthesia, surgery, and post-operative recovery.¿ ¿contraindications: the device is contraindicated whenever: the patient is known or is suspected to be allergic to materials contained in the device.Previous episodes of venous thrombosis or vascular surgical procedures at the prospective placement site.Local tissue factors will prevent proper device stabilization and/or access.¿ a lot history review (lhr) of rebs1605 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by the facility that the patient had picc placed on (b)(6) 2017, and removed on (b)(6) 2017, due to clot and break down of the skin.The picc was removed and the patient had it replaced with a powerglide.It was stated the patient may be reacting to the material in the line.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC SOLO CATHETER WITH SHERLOCK 3CG TPS STYLET 5F BASIC TRAY
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7260472
Report Number3006260740-2018-00174
Device Sequence Number11068326
Product Code LJS
UDI-Device Identifier00801741034633
UDI-Public(01)00801741034633
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K091324
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 02/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number1295118
Device Lot NumberREBS1605
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/21/2018
Initial Report FDA Received Date02/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-