• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES SPINE MNTR F/A SCREW 3.5X14; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES SPINE MNTR F/A SCREW 3.5X14; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION Back to Search Results
Catalog Number 188318314
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 12/19/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The mntr f/a screw 3.5x14 was returned to the customer quality unit (cqu) for evaluation.Visual evaluation of the mntr f/a screw 3.5x14 showed that the thread in the head of the screw driver was torn.A device history review was performed.No discrepancies were observed during the manufacturing process.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the threads on the head of the becoming torn cannot be positively determined at the moment.The likely cause of the defect is the instrument being subjected to unanticipated high amount of forces during use.This will put too much pressure on the threads on the head of the screws causing it to be torn.No issues could be identified in the manufacturing or release of this device.Therefore, this complaint will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported surgeon could not final tighten the outer nuts on top of the crosslink because the tall set screw threads had been sheared and he could not final tighten the inner crosslink set screw.Appears the screw and the inner set screw were cross threaded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MNTR F/A SCREW 3.5X14
Common Device Name
ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
MDR Report Key7260622
Report Number1526439-2018-50121
Device Sequence Number997652
Product Code NKG
UDI-Device Identifier10705034157184
UDI-Public10705034157184
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K041203
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number188318314
Device Lot Number159577
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2017
Initial Date Received by Manufacturer 02/08/2018
Initial Report FDA Received Date02/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
-
-