Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-18- user has high glucose value.Test strip (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for hi blood glucose results.The customer is concerned with test results from results obtained of hi and 166, 168, 214 mg/dl.The customer's expected fasting blood glucose test result range is 70-160 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of 592 mg/dl and hi using truemetrix meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 05/23/2018 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: (b)(6).Customer called in states the meter is reading hi.
 
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for hi blood glucose results.The customer is concerned with test results from results obtained of hi and 166, 168, 214 mg/dl.The customer's expected fasting blood glucose test result range is 70-160 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer fasting and produced test results of 592 mg/dl and hi using truemetrix meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 05/23/2018 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: (b)(6).Customer called in states the meter is reading hi.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7260948
Report Number1000113657-2018-00156
Device Sequence Number16433789
Product Code NBW
UDI-Device Identifier00096295126655
UDI-Public(01)00096295126655
Combination Product (Y/N)N
Initial Reporter StateWA
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 02/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/23/2018
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMT2242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2018
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 01/15/2018
Supplement Date Received by Manufacturer01/15/2018
Initial Report FDA Received Date02/09/2018
Supplement Report FDA Received Date02/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY; SECOND THERAPY
-
-