BAXTER HEALTHCARE - LARGO HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4474
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2017
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported problem is physical in nature and would not be found in the event logs.The sample analysis was performed and determined that the cause of the problem was the manifold assembly.The reported condition was verified.The manifold assembly will be replaced to resolve the reported issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a homechoice had an inaccurate drain and fill volume during peritoneal dialysis therapy.This event occurred during use with patient involvement.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
largo FL
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7261727
MDR Text Key99854294
Report Number1416980-2018-00680
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 02/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C4474
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2017
Initial Date Manufacturer Received 02/07/2018
Initial Date FDA Received02/09/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-