Model Number MMT-551NAP |
Device Problems
Fluid/Blood Leak (1250); Device Displays Incorrect Message (2591); Material Protrusion/Extrusion (2979)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/31/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that insulin is squirting out during the manual prime process.The customer¿s blood glucose level was 82 mg/dl.The customer was assisted with troubleshooting.Customer reports insulin squirting out during the prime process and alarmed compromised force sensor system.Customer was unable to successfully prime the set.The drive support cap is sticking out.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Pump was received with cracked end cap.Unable to confirm prime anomaly, compromised force sensor system alarm anomaly or perform the prime test due to cracked end cap.Drive support disk was inspected and no anomaly was noted.
|
|