Model Number ZXT225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017 and (b)(6) 2018.If explanted, give date: not applicable, as the lens remains implanted.However, there was a planned explant scheduled for (b)(6) 2018.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt225 23.5 diopter intraocular lens (iol), was implanted in the patient''s right eye (od) on (b)(6) 2017.Post implant the patient experienced myopic error and residual astigmatism.Post-operatively, 1-3 weeks after surgery, the patient''s refraction was -0.75 x -1.00 x 90, the uncorrected visual acuity was 20/100, and the best corrected visual acuity was 20/40.Reportedly, there was a planned explant/exchange to be performed.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 02/16/2018.Device returned to manufacturer? yes.Device evaluation: the product was received in a bottle without liquid.Visual inspection with the unaided eye revealed that the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens, further analysis is not possible.The reported complaint cannot be confirmed.Manufacturing record review: manufacturing record review of the production order and related document revealed that the product was manufactured and released according to specification.There was no discrepancy found during the review.A search revealed that no similar complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7261838
Report Number9614546-2018-00133
Device Sequence Number19122331
Product Code POE
UDI-Device Identifier05050474605862
UDI-Public(01)05050474605862(17)211101
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/01/2021
Device Model NumberZXT225
Device Catalogue NumberZXT225U235
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Initial Date Received by Manufacturer 01/12/2018
Supplement Date Received by Manufacturer04/13/2018
Initial Report FDA Received Date02/09/2018
Supplement Report FDA Received Date05/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-