PENUMBRA, INC. PX SLIM DELIVERY MICROCATHETER; DQY Back to Search Results
Model Number PXSLIM045-B
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2018-00328.
 
Event or Problem Description
The patient was undergoing a coil embolization in the unipolar iliacs using ruby coils and a px slim delivery microcatheter (px slim).It was noted that the patient's anatomy was not tortuous.During the procedure, the physician was unable to advance a ruby coil through the px slim.Therefore, the px slim and ruby coil were removed.The physician then placed a new px slim and attempted to re-advance the same ruby coil; however, the ruby coil would not advance.Subsequently, the ruby coil was removed and the procedure was completed using another ruby coil and the same px slim.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Date of report 1/16/2018.Date received by manufacturer: 1/16/2018.
 
Additional Manufacturer Narrative
Please note that the device in complaint was expected to be returned; however, additional information received from the penumbra sales representative indicated that the device was disposed of and is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.(b)(4).The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PX SLIM DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7262317
Report Number3005168196-2018-00329
Device Sequence Number19079445
Product Code DQY
UDI-Device Identifier00814548016412
UDI-Public00814548016412
Combination Product (Y/N)Y
Initial Reporter CountryBR
PMA/510(K) Number
K100826
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/16/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date09/21/2019
Device Model NumberPXSLIM045-B
Device Catalogue NumberPXSLIM045
Device Lot NumberF71716
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 01/12/2018
Supplement Date Received by Manufacturer01/16/2018
03/26/2018
Initial Report FDA Received Date02/09/2018
Supplement Report FDA Received Date05/14/2018
09/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR