• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS ELEVATE SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS ELEVATE SPINAL SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Catalog Number 8880828
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during surgery, the graft broke into two pieces as it was being implanted into the disc space.No patient complication were reported as result of this event.
 
Additional Manufacturer Narrative
Product analysis: visual review of the implant confirmed the ¿ears¿ of the -03 top component are broken off and not returned for analysis.Visual review confirms implant fracture.Witness marks and material deformation noted on the thru hole of the peek anterior ramp.These observations are consistent with overload of the implant.Due to imdrf harmonization, some previously submitted device, method, result, and conclusion codes related to this event may have been updated.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEVATE SPINAL SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7262806
Report Number1030489-2018-00201
Device Sequence Number16507887
Product Code MAX
UDI-Device Identifier00643169430815
UDI-Public00643169430815
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K172199
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/02/2025
Device Catalogue Number8880828
Device Lot Number0584994W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2018
Initial Date Received by Manufacturer 01/17/2018
Supplement Date Received by Manufacturer06/20/2018
08/08/2018
Initial Report FDA Received Date02/12/2018
Supplement Report FDA Received Date07/20/2018
10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age28 YR
-
-