| Model Number |
SENSH2428W |
| Medical Device Problem Codes |
Break (1069); Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
Perforation of Vessels (2135)
|
| Date of Event |
01/18/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Implantation of a valiant captiva stent graft system was attempted in a patient for the endovascular treatment of a 60mm diameter thoracic aortic aneurysm.Severe calcification was present, and the vessel diameter was approx.8mm.It was reported that during the index procedure, the valiant device was advanced through the sentrant sheath, but could not reach the lesion.It was noted that the strain relief was cut to insert the device delivery system but this also failed.Delivery was attempted a second time, without the use of a sheath, but was unsuccessful as the access vessel was damaged.The physician then attempted to advance the device in the contralateral artery, however this also failed, and the procedure was aborted.It was also reported that the valiant delivery system was difficult to insert through the sentrant sheath.After several attempts the device was successfully inserted, however the coating of the valiant stent graft was reportedly removed.The physician also noted an abnormality with the valve of the sheath and the valiant device during delivery.The physician commented that the device did not deliver as normal and this may have been related to the removal of the stent graft coating or due to the delivery system damage.Per the physician, the cause of the event was anatomy related, severe calcification present.No additional clinical sequelae were reported, and the patient will be monitored by their physician.Outcome attributed to adverse event: procedure was abandoned.
|
| |
|
Additional Manufacturer Narrative
|
|
Film evaluation summary: a single 3d-ct reconstruction image from an unknown study date was returned.No scale was seen on the films; therefore, no stent graft or vessel measurements could be performed.The taa was seen just caudal to the lsa.The access vessels including the iliac arteries were extensively calcified bilaterally, and both common iliacs were moderately tortuous; especially on the right side.The abdominal aorta was essentially straight; however, the thoracic aorta was severely angulated in two locations.Within the descending thoracic the aorta was acutely angulated ~105deg and the flow lumen appeared narrowed at this acute bend, and acute angulation was also seen at the arch just distal to the taa.The thoracic was calcified along the majority of its length.The exact cause of the delivery system positioning difficulties (inability to reach the lesion from either side), and the reported a ccess vessel damage, could not be determined from the single returned pre-implant image.Complete ct¿s prior to implant were not available for a more complete anatomical assessment, and images during implant showing the reported events were also not provided.It is likely that the patient¿s severely calcified and tortuous anatomy may have contributed.Multiple insertion attempts may have also led to the reported removal of the delivery system hydrocoating, as well as the difficulty to insert the device through the sentrant sheath.The cause of the reported abnormality with the sheath valve also could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: device decontaminated with cidex-opa pending further device testing to support the final product analysis findings.The device returned with the procedural sheath.The strain relief had been cut away.The device was loaded into the procedural sheath with no issues noted.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|