BOSTON SCIENTIFIC - GALWAY SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number H7493926008400
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the left anterior descending artery.A 4.00 x 8mm synergy ii drug-eluting stent was advanced to treat the lesion.However, when the device was introduced about three-fourth into the guide, the shaft of the catheter broke.The broken device together with the wire and guide was then pulled out at once.The procedure was completed using a different device.No patient complications were reported and the patient's status was good.
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the left anterior descending artery.A 4.00 x 8mm synergy ii drug-eluting stent was advanced to treat the lesion.However, when the device was introduced about three-fourth into the guide, the shaft of the catheter broke.The broken device together with the wire and guide was then pulled out at once.The procedure was completed using a different device.No patient complications were reported and the patient's status was good.
 
Additional Manufacturer Narrative
Device evaluated by mfr., eval summary attached, method codes, result codes, and conclusion codes updated.Device evaluated by mfr.: stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter was measured and was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found a hypotube break 218 mm distal from the distal end of the strain relief.Multiple hypotube kinks were also noted along the full length of the catheter.This type of damages is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the inner/outer polymer extrusion and mid-shaft section found no issues with the shaft polymer extrusion.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7263204
Report Number2134265-2018-00606
Device Sequence Number13700024
Product Code NIQ
UDI-Device Identifier08714729840015
UDI-Public08714729840015
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/07/2018
Device Model NumberH7493926008400
Device Catalogue Number39260-0840
Device Lot Number0020760788
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2018
Initial Date Received by Manufacturer 01/17/2018
Supplement Date Received by Manufacturer02/19/2018
Initial Report FDA Received Date02/12/2018
Supplement Report FDA Received Date03/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-