| Catalog Number |
383540 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported during use of the bd nexiva¿ closed iv catheter system after insertion blood was leaking from the self-sealing septum.There was no report of injury or medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: mdr level a investigation: dhr review: all challenges and samples related to the complaint were performed successfully qn review: no issues were identified relating to the complaint during production of the batch eura review: causes of the defect idenified in the eura could not be confirmed without a sample.This incident does not exceed what is considered acceptable in the eura.Sample analysis (photo): the picture showed a drop form on the back of the septum assembly.The defect was confirmed through the investigation conclusion: the root cause was not identified during the investigation.
|
| |
|