MAUDE Adverse Event Report: PENUMBRA, INC. POD PACKING COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. POD PACKING COIL; HCG, KRD Back to Search Results
Catalog Number RBYPODJ15
Medical Device Problem Codes Retraction Problem (1536); Unraveled Material (1664); Physical Resistance (2578)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the phrenic artery using pod packing coils (podjs).During the procedure, the physician successfully deployed and detached two ruby coils using a non-penumbra microcatheter.The physician then felt resistance while advancing a podj into the hub of the microcatheter, and therefore attempted to retract and remove the podj.However, the podj was unravelling while being retracted.The physician was able to completely remove the podj by retracting the pusher wire, and the procedure was then completed using a new podj and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the device in complaint was expected to be returned; however, additional information received indicated that the device was discarded by the hospital and is no longer available for return.Therefore, without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD PACKING COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7263822
Report Number3005168196-2018-00311
Device Sequence Number1222869
Product Code HCG
UDI-Device Identifier00814548016702
UDI-Public00814548016702
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K170852
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/30/2025
Device Catalogue NumberRBYPODJ15
Device Lot NumberF76856
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 01/19/2018
Supplement Date Received by Manufacturer05/24/2018
Initial Report FDA Received Date02/12/2018
Supplement Report FDA Received Date05/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
-
-