| Model Number |
VLOCA308L |
| Medical Device Problem Codes |
Break (1069); Detachment Of Device Component (1104)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
01/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter: occurred during a total laparoscopic hysterectomy procedure.The suturing device was being used while closing the vaginal cuff.The needle snapped in half after the first bite of the closure.The surgeon had took a deep bite on tissue and got hold of colpotomizer thick edge.The needle pieces fell into the cavity of the patient.It took a lot of time to locate the other half of the needle.At the end, both pieces of the needle were found and the surgeon carried on with the procedure.The surgical time was extended by thirty minutes or more due to the product problem.The incision was extended by less than one inch.In order to resolve the issue and complete the case, used another needle reload with the original suturing device and it worked fine.The patient outcome is reported as no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of three devices.Broken needle pieces were inspected under a microscope and needle was found to have been broken at suture swage.No abnormalities were observed for needle with suture thread.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all manufacturers¿ quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the bent or broken needle may occur when the needle is not loaded properly or when the needle is forced into an obstacle.This condition may also occur if excess pressure is exerted on the needle or the attached suture while the jaws are in the open position.In these situations, the needle may flex and ultimately break.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|